The macular automated photostress test.
Аннотация
PURPOSE: To introduce a standardized macular photostress test using an automated perimeter as a method to quantify macular disease severity and as a tool to distinguish optic neuropathy from macular pathology.
DESIGN
Prospective interventional pilot study.
METHODS
Twenty-five bilaterally pseudophakic subjects aged range, 65 to 84: 15 patients with varying severity of non-neovascular age-related macular degeneration (AMD), five patients with no ocular disease, and five patients with moderate primary open-angle glaucoma (POAG). Previously reported normative values served as controls for this study. Patients underwent foveal threshold testing using the Humphrey Visual Field Perimeter Model 750 (Carl Zeiss Meditec, Inc, Dublin, California, USA). Baseline measurements were compared to threshold sensitivity after photostress at one minute and then two-minute intervals until sensitivity returned to baseline. Main outcome measures were baseline foveal threshold sensitivity, foveal threshold depression, and recovery following photostress.
RESULTS
Automated macular photostress testing in macular disease (AMD) causes a decrease (P < .001) in baseline foveal sensitivity and a delay (P < .001) in recovery time to baseline sensitivity. Optic nerve pathology (POAG) does not affect (P = .343) the foveal response curve.
CONCLUSIONS
The macular automated photostress (MAP) test is an inexpensive, noninvasive, and readily accessible adjunct for evaluating patients with macular disease. This standardized protocol is useful in objectively defining disease severity, may be used to follow response to treatment, and could aid in distinguishing optic neuropathy from macular pathology.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.