MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/opo.13098

Медицинская библиотека · Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists) · 2023

Optical characterisation of two novel myopia control spectacle lenses.

Gantes-Nuñez Javier, Jaskulski Matt, López-Gil Norberto, Kollbaum Pete S

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To quantify the amount of myopic defocus, contrast modulation and other optical characteristics of two novel spectacle lenses (MiYOSMART by Hoya and Stellest by Essilor) with the inclusion of lenslets in their designs were investigated computationally and experimentally. This paper examined the hypothesis that despite the non-coaxial nature of the optics, image degradation will exist due to the fragmented nature of the base optic when imaging through the lens regions populated by lenslets.

METHODS

Optical power was evaluated by computing wavefront vergence and curvature from wavefront slope measured with the Optocraft aberrometer within 1.0 and 6.0 mm apertures across MiYOSMART hexagons and Stellest rings. Point-spread functions (PSFs) were computed using physical (wave) optics and geometrical ray optics principles, and compared with experimental measurements using a 4f optical system. Simulated retinal images and modulation transfer functions (MTFs) were computed from PSF-derived optical transfer functions (OTFs).

RESULTS

Mean lenslet power in MiYOSMART was +3.95 ± 0.10 D through the hexagons and +6.00 ± 0.15 D in Stellest in rings 1-5 and decreased by 0.42 D/ring reaching 3.50 D in the final one. Stellest lenslets included up to -0.015 microns of primary spherical aberration. PSFs and retinal images revealed simultaneous contributions of the base optic and lenslets. MTFs showed a decrease in contrast at low (1-10 c/deg) spatial frequencies (SFs) comparable to 0.25 D of defocus, and retention of diminished levels of contrast at higher SFs.

CONCLUSIONS

Varying sagittal power and consistent curvature power across the lenslets is an identifying signature of the novel non-coaxial lens design included in both spectacle lenses. Lenslet array structure itself plays a significant role in determining image characteristics. For both lenses, the blur created by the fragmented base optic contributes to the image quality. The reduced MTFs over a wide range of spatial frequencies result in lowered image contrast.

Для цитирования: Gantes-Nuñez Javier, Jaskulski Matt, López-Gil Norberto, Kollbaum Pete S. Optical characterisation of two novel myopia control spectacle lenses.. Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists). 2023. DOI: 10.1111/opo.13098.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.