DOI 10.1097/j.jcrs.0000000000001139
Comparison of intraocular lens power calculation formulas with and without total keratometry and ray tracing in patients with previous myopic SMILE.
Аннотация
PURPOSE: To evaluate the prediction error (PE) variance and absolute median PE of different intraocular lens (IOL) calculation formulas including last-generation formulas such as Barrett True-K with K, Okulix and total keratometry (TK)-based calculations with Haigis, and Barrett True-K in a simulation model in post-small-incision lenticule extraction (SMILE) eyes.
SETTINGS: Department of Ophthalmology, University Hospital Marburg, Marburg, Germany.
DESIGN
Prospective study.
METHODS
Preoperative measurements included IOL power calculation before and after SMILE surgery. The target refraction was set to be the lowest myopic refractive error in pre-SMILE eyes. The IOL power targeting at the lowest myopic refractive error in pre-SMILE eyes was selected for the post-SMILE IOL calculation of the same eye. The difference between the predicted refraction of pre- and post-SMILE eyes with the same IOL power was defined as IOL difference. The refractive change induced by SMILE was defined as the difference between preoperative and postoperative manifest refraction.
RESULTS
98 eyes from 49 patients underwent bilateral myopic SMILE. The PE variance of Okulix was not significantly different compared with Barrett True-K with TK ( P = .471). The SDs of the mean PEs were ±0.413 D (Haigis-TK), ±0.453 D (Okulix), ±0.471 D (Barrett True-K with TK), ±0.556 D (Haigis-L), and ±0.576 D (Barrett True-K with K). The mean absolute PE was 0.340 D, 0.353 D, 0.404 D, 0.511 D, and 0.715 D for Haigis-TK, Okulix, Barrett True-K with TK, Barrett True-K with K, and Haigis-L, respectively. The highest percentage of eyes within ±0.50 D was achieved by Okulix, followed by Haigis-TK, Barrett True-K with TK, Barrett True-K with K, and Haigis-L.
CONCLUSIONS
Results suggest that Haigis in combination with TK, Okulix, and Barrett True-K with and without TK offer good options for accurate IOL power calculation after SMILE. Haigis-L showed a tendency for myopic shift in eyes after previous SMILE.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.