Respiratory care · 2021

Agreement Between Transcutaneous Monitoring and Arterial Blood Gases During COPD Exacerbation.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Sørensen Kasper M, Leicht Rebecca V, Carlsson Christian J, Elvekjaer Mikkel, Porsbjerg Celeste, Aasvang Eske K, Meyhoff Christian S

Аннотация

BACKGROUND: Transcutaneous measurements of CO2 and O2 ([Formula: see text], [Formula: see text]) are noninvasive and allow for continuous monitoring in adults with exacerbation of COPD, but substantial accuracy issues may exist. We investigated agreement between results of arterial blood gas analysis and transcutaneous measurements of CO2 and O2 in patients with COPD.

METHODS: Adult subjects were monitored after acute admission to a respiratory intermediate care unit or ICU due to exacerbation of COPD and with ongoing noninvasive ventilation or immediately following extubation. Monitored variables were continuous transcutaneous measurement and simultaneous routine arterial blood gas analysis. Agreement between measurements was assessed by calculating bias with 95% limits of agreement for single-point estimates of [Formula: see text] versus [Formula: see text] and versus [Formula: see text], and for changes in transcutaneous measurements between 2 time points ([Formula: see text] and [Formula: see text]). We considered limits of agreement within ± 7.5 mm Hg to be acceptable.

RESULTS: A total of 57 transcutaneous measurements were made in 20 subjects for comparison with concurrent arterial blood gas analysis at 36 time points. The bias (limits of agreement) for [Formula: see text] and [Formula: see text] was 2.5 mm Hg (-10.6 to 15.6 mm Hg) and 11.2 mm Hg (-28.2 to 50.6 mm Hg), respectively. The bias for [Formula: see text] and [Formula: see text] was 2.3 mm Hg (-3.8 to 8.3 mm Hg) and -5.3 mm Hg (-37.5 to 27 mm Hg), respectively.

CONCLUSIONS: [Formula: see text] and [Formula: see text] did not accurately reflect results from arterial blood gas analyses in this study of mostly hypercapnic subjects. Agreement between changes in CO2 during the monitoring period was acceptable, however, and transcutaneous monitoring may be used for continuous monitoring of [Formula: see text] in conjunction with arterial blood gas analysis for reference.