Ethyl glucuronide and ethyl sulphate determination in serum by liquid chromatography-electrospray tandem mass spectrometry.
СтатьяВалидационное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Morini Luca, Politi Lucia, Zucchella Alessandra, Polettini Aldo
Аннотация
BACKGROUND: Ethyl glucuronide (EtG) and ethyl sulphate (EtS) are two ethanol metabolites that can be detected in serum up to 8 h after ethanol elimination. Their presence is therefore indicative of recent ethanol consumption in case of delayed sampling after an event (e.g. car crash).
METHODS: A LC-ESI-MS-MS method for the determination of EtG and EtS in serum was developed and validated. Two product ions together with the parent ions were monitored for identification. Pentadeuterated-EtG was used as internal standard.
RESULTS: Excellent linearity for EtG (from 0.22 to 45 micromol/l) and EtS (from 0.40 to 80 micromol/l) was observed (r(2)>or=0.9998). LOD and LLOQ were 0.04 and 0.20 micromol/l for EtG and 0.08 and 0.40 micromol/l for EtS, respectively. Accuracy (bias) and precision (relative standard deviation), studied at four different quality control levels, were always better than 7%. Matrix effects were found to be negligible. The method was applied to several samples obtained from known alcoholics and social drinkers.
CONCLUSIONS: A sensitive and specific determination of EtG and EtS in serum samples was achieved despite a simple and fast sample preparation. To our knowledge, this is the first fully validated method for the simultaneous determination of the two alcohol metabolites in serum.
METHODS: A LC-ESI-MS-MS method for the determination of EtG and EtS in serum was developed and validated. Two product ions together with the parent ions were monitored for identification. Pentadeuterated-EtG was used as internal standard.
RESULTS: Excellent linearity for EtG (from 0.22 to 45 micromol/l) and EtS (from 0.40 to 80 micromol/l) was observed (r(2)>or=0.9998). LOD and LLOQ were 0.04 and 0.20 micromol/l for EtG and 0.08 and 0.40 micromol/l for EtS, respectively. Accuracy (bias) and precision (relative standard deviation), studied at four different quality control levels, were always better than 7%. Matrix effects were found to be negligible. The method was applied to several samples obtained from known alcoholics and social drinkers.
CONCLUSIONS: A sensitive and specific determination of EtG and EtS in serum samples was achieved despite a simple and fast sample preparation. To our knowledge, this is the first fully validated method for the simultaneous determination of the two alcohol metabolites in serum.