DOI 10.1016/j.jcrs.2012.12.031

Медицинская библиотека · Journal of cataract and refractive surgery · 2013

Scheimpflug analysis of corneal power changes after myopic excimer laser surgery.

Savini Giacomo, Hoffer Kenneth J, Carbonelli Michele, Barboni Piero

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To assess the ability of corneal power measurements by a rotating Scheimpflug camera to measure the refractive change induced by myopic excimer laser surgery.

SETTING

G.B. Bietti Foundation-IRCCS, Rome, Italy.

DESIGN

Evaluation of diagnostic test.

METHODS

The following corneal power measurements by the Pentacam Scheimpflug camera were analyzed: average keratometry (K), true net power (calculated by Gaussian optics formula), and total corneal refractive power (TCRP) at 2.0 mm, 3.0 mm, and 4.0 mm, calculated by ray tracing on a ring and as the average of the zone inside the ring. The difference between the preoperative and postoperative values was compared with the subjective surgically induced refractive change (SIRC) and with the difference between the preoperative and the postoperative anterior corneal power measured by Placido corneal topography (Keratron).

RESULTS

In 36 consecutive eyes, the average K significantly underestimated the SIRC as determined by subjective refraction (-4.47 diopters [D] ± 1.81 [SD]) and corneal topography (-4.38 ± 1.81 D). The 3.0 mm and 4.0 mm ring total corneal refractive power significantly overestimated the SIRC. The remaining values did not show statistically significant differences with respect to the SIRC. The 3.0 mm zone TCRP and the 2.0 mm ring TCRP provided the lowest median difference compared with the SIRC (-0.07 D and -0.17 D, respectively) and the closest agreement.

CONCLUSIONS

The corneal power values provided by the Scheimpflug camera accurately reflected the SIRC after myopic excimer laser surgery. The best options seem to be the 3.0 mm zone TCRP and the 2.0 mm ring TCRP.

Для цитирования: Savini Giacomo, Hoffer Kenneth J, Carbonelli Michele, Barboni Piero. Scheimpflug analysis of corneal power changes after myopic excimer laser surgery.. Journal of cataract and refractive surgery. 2013. DOI: 10.1016/j.jcrs.2012.12.031.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.