Predictability of corneal flap thickness in laser in situ keratomileusis using a 200 kHz femtosecond laser.
Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Cummings Arthur B, Cummings Brendan K, Kelly Gabrielle E
Аннотация
PURPOSE: To compare the intended versus the resultant thickness of laser in situ keratomileusis (LASIK) flaps created with a new femtosecond laser and to report the 3-month postoperative refractive outcomes.
SETTING: Wellington Eye Clinic, Dublin, Ireland.
DESIGN: Retrospective consecutive case series.
METHODS: A consecutive series of eyes having LASIK flaps created using a Wavelight FS200 femtosecond laser were included in this analysis. Eyes with preoperative spherical equivalent (SE) myopic refractive errors up to -12.00 diopters (D), hyperopic refractive errors up to +5.50 D, and astigmatism up to 6.00 D received primary LASIK. Outcomes included postoperative flap thickness and 3-month postoperative refraction.
RESULTS: Outcomes in 431 eyes of 258 patients having LASIK with femtosecond flap creation were assessed. With an intended flap thickness of 120 μm, the mean post-LASIK flap thickness was 120.23 μm ± 13.94 (SD). In 349 of 813 eyes, EX500 optical pachymetry measurements were compared with Scheimpflug pachymetry readings. The mean corneal thickness was 553.52 ± 29.81 μm with optical pachymetry and 553.39 ± 29.12 μm with the Pentacam rotating Scheimpflug camera. In 162 eyes, the flap thickness was measured with optical coherence tomography; the mean was 121.94 ± 10.52 μm. Achieved flap dimensions were as intended.
CONCLUSIONS: This study provides evidence of the predictability and lower variability of LASIK flaps created using the femtosecond laser. Consecutive patients having LASIK had good, predictable, and stable outcomes at the 3-month mark.
SETTING: Wellington Eye Clinic, Dublin, Ireland.
DESIGN: Retrospective consecutive case series.
METHODS: A consecutive series of eyes having LASIK flaps created using a Wavelight FS200 femtosecond laser were included in this analysis. Eyes with preoperative spherical equivalent (SE) myopic refractive errors up to -12.00 diopters (D), hyperopic refractive errors up to +5.50 D, and astigmatism up to 6.00 D received primary LASIK. Outcomes included postoperative flap thickness and 3-month postoperative refraction.
RESULTS: Outcomes in 431 eyes of 258 patients having LASIK with femtosecond flap creation were assessed. With an intended flap thickness of 120 μm, the mean post-LASIK flap thickness was 120.23 μm ± 13.94 (SD). In 349 of 813 eyes, EX500 optical pachymetry measurements were compared with Scheimpflug pachymetry readings. The mean corneal thickness was 553.52 ± 29.81 μm with optical pachymetry and 553.39 ± 29.12 μm with the Pentacam rotating Scheimpflug camera. In 162 eyes, the flap thickness was measured with optical coherence tomography; the mean was 121.94 ± 10.52 μm. Achieved flap dimensions were as intended.
CONCLUSIONS: This study provides evidence of the predictability and lower variability of LASIK flaps created using the femtosecond laser. Consecutive patients having LASIK had good, predictable, and stable outcomes at the 3-month mark.