[AcrySof® Toric (SN60T) intraocular lens implantation: refractive predictibility and aberrometric impact of decentration].
Оценочное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Humbert G, Colin J, Touboul D
Аннотация
PURPOSE: To evaluate refractive outcomes of toric intraocular lens (IOL) implantation with a detailed analysis of decentration and its effect on aberrometry.
METHODS: This retrospective study enrolled 19 eyes implanted with SN60T AcrySof(®) Toric (Alcon - USA) IOL's. Spherocylindric correction was studied in depth by the Alpins method, and retro-illumination images were used to analyze the alignment of the IOL with its intended axis. IOL decentration as well as its aberrometric impact were evaluated with a strict, novel protocol.
RESULTS: Three months postoperatively, uncorrected distance visual acuity was greater or equal to 20/40 in 94.74% of cases, postoperative subjective cylinder was less or equal to 0.5 D in 68.42% of eyes, with a mean index of success of 0.24. Mean error of toric IOL alignment was 5.68° (0 to 14). Mean IOL decentration was 0.78 mm (0 to 1.78) with a mean coma and trefoil of 0.18 μ (0.06 to 0.33) and 0.19 μ (0.05 to 0.51), respectively. The larger the IOL decentration, the higher the optical aberrations were.
CONCLUSION: Toric intraocular lens implantation is an effective, safe and predictable method of spherocylindrical correction during cataract surgery, with a refractive accuracy similar to that of LASIK in the treatment of astigmatism in young patients. IOL decentration produces optical aberrations including coma and trefoil, which interfere with visual performance.
METHODS: This retrospective study enrolled 19 eyes implanted with SN60T AcrySof(®) Toric (Alcon - USA) IOL's. Spherocylindric correction was studied in depth by the Alpins method, and retro-illumination images were used to analyze the alignment of the IOL with its intended axis. IOL decentration as well as its aberrometric impact were evaluated with a strict, novel protocol.
RESULTS: Three months postoperatively, uncorrected distance visual acuity was greater or equal to 20/40 in 94.74% of cases, postoperative subjective cylinder was less or equal to 0.5 D in 68.42% of eyes, with a mean index of success of 0.24. Mean error of toric IOL alignment was 5.68° (0 to 14). Mean IOL decentration was 0.78 mm (0 to 1.78) with a mean coma and trefoil of 0.18 μ (0.06 to 0.33) and 0.19 μ (0.05 to 0.51), respectively. The larger the IOL decentration, the higher the optical aberrations were.
CONCLUSION: Toric intraocular lens implantation is an effective, safe and predictable method of spherocylindrical correction during cataract surgery, with a refractive accuracy similar to that of LASIK in the treatment of astigmatism in young patients. IOL decentration produces optical aberrations including coma and trefoil, which interfere with visual performance.