Comparison of Tono-Pen AVIA intraocular pressure measurements performed at limbus with central corneal Tono-Pen AVIA intraocular pressure.

Sharma Reetika, Majumdar Saptorishi, Sobti Amit, Arora Tarun, Agarwal Tushar, Dada Tanuj

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare Tono-Pen AVIA (TPA) intraocular pressure (IOP) measurements performed at the limbus with central corneal Tono-Pen AVIA (TPA) intraocular pressure.

METHODS

Fifty-one patients attending our outpatient department with a best corrected visual acuity of ≥ 20/40, no corneal disease, and no astigmatism greater than 1.5 diopters underwent routine ophthalmological examination. Patients with ametropia greater than or equal to ± 5 diopters, a history of recent contact lens wear, and central corneal thickness of <480 µm or >550 µm were excluded from the study. IOP was recorded first in the corneal center, followed by tonometry measurements obtained from the temporal, nasal, superior, and inferior limbus using TPA. IOP measurements were compared using the paired t test. Correlation was derived by Pearson correlation coefficient, and regression equations were derived.

RESULTS

The mean age of patients in our study was 43.86 ± 15.39 years. The mean TPA IOP from the central cornea was 16.59 ± 3.75 mm Hg. There was a statistically significant difference between Tono-Pen IOP from the central cornea and that recorded from all 4 limbal positions (P < 0.001). There was a positive correlation between the central and the limbal IOP obtained with Tono-Pen. Correlation was maximum for the temporal limbus.

CONCLUSIONS

Limbal TPA IOP is significantly higher than central TPA IOP. There remains a good correlation between central and limbal IOP. This was found to be greatest for the temporal limbus.

Для цитирования: Sharma Reetika, Majumdar Saptorishi, Sobti Amit, Arora Tarun, Agarwal Tushar, Dada Tanuj. Comparison of Tono-Pen AVIA intraocular pressure measurements performed at limbus with central corneal Tono-Pen AVIA intraocular pressure.. Cornea. 2013. DOI: 10.1097/ico.0b013e31827c9d25.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.