Journal of cataract and refractive surgery · 2012

Comparability and repeatability of corneal astigmatism measurements using different measurement technologies.

Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Visser Nienke, Berendschot Tos T J M, Verbakel Frenne, de Brabander John, Nuijts Rudy M M A

Аннотация

PURPOSE: To determine the comparability and repeatability of corneal astigmatism measurements obtained with different devices and determine the interobserver variability of a new automated keratometer.

SETTING: University Eye Clinic Maastricht, the Netherlands.

DESIGN: Prospective cohort study.

METHODS: The right eye of healthy subjects was examined with the following 6 devices: IOLMaster (automated keratometry), Lenstar (automated keratometry), SMI Reference Unit 3 (automated keratometry), Javal (manual keratometry), KR-1W (corneal topography), and Pentacam (Scheimpflug imaging). An experienced operator obtained 3 repeated measurements. An inexperienced operator obtained additional measurements with the SMI Reference Unit 3. Astigmatism vector analysis was used to determine the comparability, repeatability, and interobserver variability.

RESULTS: Corneal astigmatism vectors measured by automated, manual, or simulated keratometry were comparable except for the Pentacam equivalent keratometry (K) (P<.001, repeated-measures analysis of variance [ANOVA]). The mean difference between the equivalent K and other K values was 0.18 to 0.29 diopter (D) (P<.05, Hotelling trace multivariate ANOVA). The mean differences between automated, manual, and simulated keratometry were small (≤0.12 D). The within-subject standard deviation ranged from 0.05 D @ 21 degrees (KR-1W) to 0.18 D @ 23 degrees (Lenstar). The SMI Reference Unit showed small mean differences and comparable repeatability between the experienced operator and the inexperienced operator.

CONCLUSIONS: Vector analysis showed comparable corneal astigmatism measurements using automated, manual, and simulated keratometry. Pentacam equivalent K values were not comparable with those of the other keratometers. The repeatability of astigmatism magnitudes was acceptable; however, the repeatability of astigmatism meridians was moderate. The SMI Reference Unit showed good interobserver variability.