MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2147/opth.s29660

Медицинская библиотека · Clinical ophthalmology (Auckland, N.Z.) · 2012

Comparison of outcomes of conventional WaveLight(®) Allegretto Wave(®) and Technolas(®) excimer lasers in myopic laser in situ keratomileusis.

Han Daphne Cy, Chen Jean, Htoon Hla Myint, Tan Donald Th, Mehta Jodhbir S

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To compare the results of laser in situ keratomileusis for myopia using WaveLight(®) Allegretto Wave(®) Eye-Q(®) and Technolas(®) 217z excimer lasers.

METHOD: A retrospective, comparative case series of 442 eyes matched for age and myopia: half each were treated with Allegretto's wavefront-optimized algorithm and Technolas PlanoScan. Outcome measures were postoperative mean logarithm of the minimum angle of resolution (logMAR) uncorrected visual acuity (UCVA), manifest refraction spherical equivalent (MRSE), cylinder, safety and efficacy indices, refractive predictability, and optical zone size selection. Refractive predictability of a subgroup treated for -2.50 to -4.0 diopter (D) was analyzed separately.

RESULTS

At mean follow-up of 80.5 days, mean logMAR UCVA, mean MRSE and mean postoperative cylinder were 0.02 ± 0.07 (range -0.12 to 0.30), 0.27 ± 0.36 D (range -1.25 to 1.50 D) and -0.33 ± 0.30 D (range 0.00 to -1.50 D) for Allegretto versus 0.02 ± 0.08 (range -0.12 to 0.40), 0.095 ± 0.47 D (range -1.25 to 1.13 D) and -0.44 ± 0.5 2 D (range 0.00 to -2.25 D) for Technolas (P = 0.98, 0.80 and 0.006). Mean safety and efficacy indices were 1.05 ± 0.13 (0.75-1.33) and 0.97 ± 0.13 (0.50-1.33) for Allegretto and 1.07 ± 0.14 (0.75-1.49) and 0.97 ± 0.17 (0.40-1.49) for Technolas (P = 0.23 and 0.69). Proportions of eyes achieving postoperative MRSE within ±1.0 D, ±0.5 D, and ±0.25 D were 98.2%, 91.9% and 75.6% for Allegretto and 99.1%, 97.8% and 72.4% for Technolas (P = 0.68, 0.20 and 0.51). Mean optical zone size selected was 6.48 ± 0.10 mm (range 6.0-6.5 mm) for Allegretto and 6.38 ± 0.19 mm (range 5.6-6.6 mm) for Technolas (P < 0.001). Of the subgroup with treatment between -2.5 and -4.0 D, 86.8% and 58.5% of eyes treated with Allegretto achieved postoperative MRSE within ±0.50 D and ±0.25 D versus 70.4% and 44.4% for Technolas (P = 0.006 and 0.057).

CONCLUSION

No differences were seen in postoperative mean logMAR UCVA, MRSE, safety and efficacy indices between the two lasers. Allegretto produced less residual astigmatism, possibly improved refractive predictability, and required smaller optical zone selection.

Для цитирования: Han Daphne Cy, Chen Jean, Htoon Hla Myint, Tan Donald Th, Mehta Jodhbir S. Comparison of outcomes of conventional WaveLight(®) Allegretto Wave(®) and Technolas(®) excimer lasers in myopic laser in situ keratomileusis.. Clinical ophthalmology (Auckland, N.Z.). 2012. DOI: 10.2147/opth.s29660.