Journal of cataract and refractive surgery · 2012

Univariate and bivariate polar value analysis of corneal astigmatism measurements obtained with 6 instruments.

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Lee Hun, Chung Jae Lim, Kim Eung Kweon, Sgrignoli Bradford, Kim Tae-im

Аннотация

PURPOSE: To compare the corneal astigmatism measurements from 6 instruments in preoperative assessment for toric intraocular lens (IOL) implantation.

SETTING: Institute of Vision Research, Department of Ophthalmology, Yonsei University College of Medicine, Seoul, South Korea.

DESIGN: Prospective comparative observational study.

METHODS: This study included patients with cataract and more than 1.00 diopter (D) of corneal astigmatism. For preoperative evaluation of toric IOL implantation, the net astigmatism was evaluated using manual keratometry, autokeratometry, partial coherence interferometry (PCI) (IOLMaster), corneal topography/ray-tracing aberrometry (iTrace), scanning-slit topography (Orbscan), and Scheimpflug imaging (Pentacam). All net astigmatisms were converted to polar values. Using the astigmatism measurements from manual keratometry as a standard, Bland-Altman analysis, linear mixed-model, and bivariate graphic analysis were performed.

RESULTS: The study group comprised 257 eyes of 141 patients. Bland-Altman plots showed good agreement between manual keratometry and each instrument for polar values. There was no significant between-instrument difference in KP(90) and KP(135) in the linear mixed model analysis or in bivariate polar values in bivariate confidence ellipses.

CONCLUSION: The corneal astigmatism measurements from autokeratometry, PCI, corneal topography/ray-tracing aberrometry, scanning-slit topography, and Scheimpflug imaging were comparable to those from manual keratometry and can be used interchangeably with manual keratometry to measure corneal astigmatism.