Univariate and bivariate polar value analysis of corneal astigmatism measurements obtained with 6 instruments.

Lee Hun, Chung Jae Lim, Kim Eung Kweon, Sgrignoli Bradford, Kim Tae-im

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the corneal astigmatism measurements from 6 instruments in preoperative assessment for toric intraocular lens (IOL) implantation.

SETTING

Institute of Vision Research, Department of Ophthalmology, Yonsei University College of Medicine, Seoul, South Korea.

DESIGN

Prospective comparative observational study.

METHODS

This study included patients with cataract and more than 1.00 diopter (D) of corneal astigmatism. For preoperative evaluation of toric IOL implantation, the net astigmatism was evaluated using manual keratometry, autokeratometry, partial coherence interferometry (PCI) (IOLMaster), corneal topography/ray-tracing aberrometry (iTrace), scanning-slit topography (Orbscan), and Scheimpflug imaging (Pentacam). All net astigmatisms were converted to polar values. Using the astigmatism measurements from manual keratometry as a standard, Bland-Altman analysis, linear mixed-model, and bivariate graphic analysis were performed.

RESULTS

The study group comprised 257 eyes of 141 patients. Bland-Altman plots showed good agreement between manual keratometry and each instrument for polar values. There was no significant between-instrument difference in KP(90) and KP(135) in the linear mixed model analysis or in bivariate polar values in bivariate confidence ellipses.

CONCLUSION

The corneal astigmatism measurements from autokeratometry, PCI, corneal topography/ray-tracing aberrometry, scanning-slit topography, and Scheimpflug imaging were comparable to those from manual keratometry and can be used interchangeably with manual keratometry to measure corneal astigmatism.

Для цитирования: Lee Hun, Chung Jae Lim, Kim Eung Kweon, Sgrignoli Bradford, Kim Tae-im. Univariate and bivariate polar value analysis of corneal astigmatism measurements obtained with 6 instruments.. Journal of cataract and refractive surgery. 2012. DOI: 10.1016/j.jcrs.2012.04.035.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.