DOI 10.1097/j.jcrs.0000000000001380

Медицинская библиотека · Journal of cataract and refractive surgery · 2024

Precision of a new SS-OCT biometer to measure anterior segment parameters and agreement with 3 instruments with different measurement principles.

Venkataraman Abinaya Priya, Domínguez-Vicent Alberto, Selin Paulina, Brautaset Rune, Montés-Micó Robert

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the repeatability of a new swept source optical coherence tomography (SS-OCT)-based biometer to measure anterior segment parameters and to assess the agreement with 3 other imaging devices based on different measurement principles.

SETTING

Unit of Eye and Vision, Department of Clinical Neuroscience, Karolinska Institute, Stockholm, Sweden.

DESIGN

Prospective, comparative case series.

METHODS

3 consecutive measurements were obtained in unoperated eyes with the Eyestar900 (SS-OCT), Lenstar 900, MS-39, and Sirius. The following anterior segment parameters were evaluated: central corneal thickness (CCT), corneal diameter (CD), aqueous depth (AQD), and corneal power metrics. The repeatability limit (Rlim), coefficient of variation (CoV), and a repeated measures Bland-Altman analysis were performed.

RESULTS

74 eyes of 74 participants were measured. The Rlims for CCT, CD, and AQD were lower than 10 μm, 0.3 mm, and 0.10 mm for all devices, respectively. The corresponding CoVs for these parameters never exceeded 1.2%. The Rlim for the corneal power metrics never exceeded 0.60 diopter (D) for any of the instruments. Lenstar showed the best agreement with the MS-39 to measure CCT, CD, and AQD (limit of agreement interval, LoA: 15.54 μm, 0.55 mm, and 0.16 mm, respectively). The mean difference for keratometry parameters was lower than 0.3 D for all device comparisons, and the LoA interval ranged between 0.52 D and 1.21 D.

CONCLUSIONS

The repeatability for measuring anterior segment parameters was good, and the agreement among all the instruments was good for CD and AQD measurements. However, for CCT and keratometer parameters, the instruments cannot be used interchangeably due to large LoA interval.

Для цитирования: Venkataraman Abinaya Priya, Domínguez-Vicent Alberto, Selin Paulina, Brautaset Rune, Montés-Micó Robert. Precision of a new SS-OCT biometer to measure anterior segment parameters and agreement with 3 instruments with different measurement principles.. Journal of cataract and refractive surgery. 2024. DOI: 10.1097/j.jcrs.0000000000001380.