DOI 10.1097/j.jcrs.0000000000001367
Attaining the optimal flange technique for transscleral capsular bag stabilization using iris hooks.
Аннотация
PURPOSE: To investigate the flange properties of different iris hooks.
SETTING
Vienna Institute for Research in Ocular Surgery (VIROS), Hanusch Hospital, Vienna, Austria.
DESIGN
Laboratory study.
METHODS
The flanging properties of 4 different iris hooks made from polypropylene (PP), elastic polymer (EP), and nylon were investigated with different heating distances and both with and without forceps gripping. The maximum diameter of the flanges was measured, and the shape of the flanges was evaluated.
RESULTS
Although both nylon and EP iris hooks had too small flange diameters for intrascleral fixation, PP iris hooks had a sufficient flange diameter (>330 μm) and mushroom-like shape. Furthermore, in PP hooks, heating distance was directly proportional to flange diameter.
CONCLUSIONS
The findings of this study suggest that only PP iris hooks are suitable for flanged intrascleral fixation, which is off-label, to secure adequate fixation.