Comparison of 4 point-of-care techniques to detect correct positioning of nasogastric tubes in dogs (2020-2021).

Ramesh Meera, Brooks Aimee C, Thomovsky Elizabeth, Johnson Paula, Fulkerson Carrie, Parnell Nollie, Weng Hsin-Yi

Observational Study, VeterinaryСтатьяПеревод в очереди

Аннотация

OBJECTIVE

To compare 4 point-of-care (POC) techniques to assess nasogastric (NG) tube placement versus radiographs as a reference standard. POC methods included air inflation with auscultation, fluid aspiration with pH measurement, ultrasonography, and capnography.

DESIGN

Prospective observational study in hospitalized dogs between 2020 and 2021.

SETTING

University teaching hospital.

ANIMALS: Fifty-one dogs requiring NG tube placement as part of their normal care.

INTERVENTIONS: After standard blind NG tube placement, each POC method was performed following standardized instructions. All POC methods were scored as to whether the investigator believed the tube to be in the gastrointestinal tract (as indicated by positive auscultation of borborygmus during insufflation, positive fluid aspiration with pH ≤5, presence of hyperechoic shadow in the esophagus, or absence of capnographic waveform). Subsequently, radiographs were taken to determine NG tube position as a gold standard. The sensitivity, specificity, and accuracy of each test as compared to 2-view thoracic radiographs were determined.

MEASUREMENTS AND MAIN

RESULTS

Sensitivity, specificity, and accuracy for each POC technique were as follows: air auscultation (84.4%, 50.5%, and 80.4%, respectively), neck ultrasound (95.6%, 83.3%, and 94.1%, respectively), capnography (91.1%, 33.3%, and 84.3%, respectively), and fluid aspiration with pH measurement (22.2%, 100%, and 31.4%, respectively).

CONCLUSIONS

Among the 4 techniques evaluated, neck ultrasound had the best overall performance for assessing NG tube placement. Fluid aspiration with pH measurement might also have potential due to perfect specificity, but its clinical utility may be limited by low sensitivity and accuracy. Nonetheless, 2-view thoracic radiography should still be considered the standard method for confirmation of NG tube placement as none of the 4 POC techniques investigated showed both high sensitivity and perfect specificity.

Для цитирования: Ramesh Meera, Brooks Aimee C, Thomovsky Elizabeth, Johnson Paula, Fulkerson Carrie, Parnell Nollie, Weng Hsin-Yi. Comparison of 4 point-of-care techniques to detect correct positioning of nasogastric tubes in dogs (2020-2021).. Journal of veterinary emergency and critical care (San Antonio, Tex. : 2001). 2023. DOI: 10.1111/vec.13327.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.