PURPOSE: To compare 5-year corneal endothelial safety of 3 minimally invasive glaucoma surgery (MIGS) devices (iStent inject , Hydrus Microstent, CyPass Micro-Stent).

SETTING

U.S. multicenter trials.

DESIGN

Post hoc 5-year analysis from prospective randomized single-masked pivotal trials.

METHODS

Mild to moderate open-angle glaucoma subjects received a MIGS implant with phacoemulsification (implant + phaco) or phaco alone (control). In addition, 5-year end points comparing the implant and control groups included proportion of eyes with significant endothelial cell loss (ECL) (>30% or ≥30% vs baseline) and mean endothelial cell density (ECD).

RESULTS

Comparable proportions of eyes in the iStent inject + phaco and control groups had significant 60-month ECL (9.4% vs 6.3%, respectively, diff: 3.2%, 95% CI, -5.0% to 11.3%, P = .77). Hydrus (20.8% vs 10.6%, diff: 10.2%, 95% CI, 3.2% to 17.2%, P = .01) and CyPass (27.2% vs 10.0%, diff: 17.2%, 95% CI, 5.6% to 28.7%, P = .02) had more eyes with ECL vs controls (iStent inject 1.49X, Hydrus 1.96X, CyPass 2.72X vs controls). Mean ECD over 60 months for iStent inject was indistinguishable vs control, whereas greater ECL was observed primarily 3 months postoperative (Hydrus) or accelerated after 2 years (CyPass). No iStent inject or Hydrus subjects developed persistent corneal edema, whereas 7 CyPass-implanted eyes developed ECL-related complications.

CONCLUSIONS

Through 5 years postoperative, there were no differences in proportion of eyes with significant ECL or mean ECD between the iStent inject and control groups. There was greater 5-year ECL and lower ECD in the Hydrus and CyPass groups vs controls. The Hydrus ECL rate mirrored control after 3 months; the CyPass ECL rate accelerated vs control.

Для цитирования: Ahmed Iqbal Ike K, Sheybani Arsham, De Francesco Ticiana, Samuelson Thomas W. Corneal endothelial safety profile in minimally invasive glaucoma surgery.. Journal of cataract and refractive surgery. 2024. DOI: 10.1097/j.jcrs.0000000000001365.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.