Influence of a Capsular Tension Ring on Capsular Bag Behavior of a Plate Haptic Intraocular Lens: An Intraindividual Randomized Trial.

Schartmüller Daniel, Röggla Veronika, Schwarzenbacher Luca, Meyer Elias Laurin, Abela-Formanek Claudette, Leydolt Christina, Menapace Rupert

Рандомизированное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the influence of a capsular tension ring (CTR) on rotational stability, decentration, tilt, and axial stability of an 11.0-mm plate haptic intraocular lens (IOL).

DESIGN

Intraindividual, randomized, double-masked, controlled clinical trial.

PARTICIPANTS

Patients scheduled for sequential same-day bilateral cataract surgery.

METHODS

All patients were randomized to receive a CTR and a plate haptic IOL in one eye and a plate haptic IOL in the fellow eye only. Intraocular lens axis assessment was performed at the end of surgery, 1 hour, 1 week, 1 month, and 6 months using a high-precision evaluation method. Decentration and tilt of the crystalline and pseudophakic lenses were assessed before surgery and at 1 week and 6 months using an anterior segment OCT.

MAIN OUTCOME MEASURES: Rotational stability from the end of surgery to 6 months and at all follow-up visits, decentration and tilt at 6 months, and differences in axial shift between 1 week and 6 months.

RESULTS

One hundred thirty eyes of 65 patients were included in the study. Absolute rotation from the end of surgery to 6 months was 2.8 ± 3.9° and 3.2 ± 5.3° for the CTR and control groups, respectively (P = 0.613). Intraocular lens decentration and IOL tilt at 6 months were 0.29 ± 0.1 mm and 0.24 ± 0.1 mm and 6.7 ± 2.8° and 5.6 ± 1.6° for the CTR and control groups, respectively (P = 0.058; P < 0.01). A posterior IOL shift of 0.31 ± 0.31 mm and 0.19 ± 0.14 mm was observed in the CTR and control groups, respectively.

CONCLUSIONS

Concomitant implantation of a CTR and a plate haptic IOL did not improve the overall rotational stability of the IOL compared with the control group. Against expectations, higher values of decentration, tilt, and axial shift were observed in the CTR group. The simultaneous use of a CTR and a plate haptic IOL in the absence of zonular weakness at the time of cataract surgery should be considered with caution.

FINANCIAL DISCLOSURE(S): Proprietary or commercial disclosure may be found in the Footnotes and Disclosures at the end of this article.

Для цитирования: Schartmüller Daniel, Röggla Veronika, Schwarzenbacher Luca, Meyer Elias Laurin, Abela-Formanek Claudette, Leydolt Christina, Menapace Rupert. Influence of a Capsular Tension Ring on Capsular Bag Behavior of a Plate Haptic Intraocular Lens: An Intraindividual Randomized Trial.. Ophthalmology. 2023. DOI: 10.1016/j.ophtha.2023.10.031.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.