Stand-alone mapping using different transluminal mapping catheters--an accurate and safe way to isolate all pulmonary veins with the cryoballoon?
Аннотация
PURPOSE: Cryoballoon isolation of the pulmonary veins (PVI) is an effective and safe method in the treatment of patients with paroxysmal atrial fibrillation (PAF). The circular mapping catheter Achieve® (Medtronic Inc., Minneapolis, USA) can be introduced into the pulmonary vein (PV) through the inner lumen of the balloon catheter, allowing online mapping of the PV electrograms during ablation. We prospectively compared the accuracy of this catheter in both available sizes to a standard circular mapping catheter (SCMC) in order to determine bidirectional PV block.
METHODS
Patients with symptomatic PAF underwent cryoballoon ablation using either the Achieve® 15-mm catheter (group 1, 15 patients) or the Achieve® 20 mm (group 2, 15 patients). PV potentials were recorded using Achieve® before, during, and after ablation, and exit block was obtained by pacing from inside the PV. Accuracy of PV potential detection was controlled by a SCMC before and after ablation.
RESULTS
Rate of PV which could be isolated exclusively using the Achieve® as guidewire was 98 % (59/60) in group 1 and 93 % (57/60) in group 2. Online signal recording during ablation was possible in 40 and 60 % of PV (p = 0.037), respectively. Final Achieve® diagnosis was accurate in 55/60 (92 %) of the PVs and 12/15 (80 %) of patients in group 1 and 60/60 (100 %) of PV and 15/15 (100 %) of patients in group 2.
CONCLUSION
Stand-alone mapping using Achieve® 15 mm resulted in a significant lack of diagnostic accuracy. Achieve® 20 mm provided excellent diagnostic accuracy comparable to a SCMC and should be preferentially used.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.