Effects of hybrid closed-loop system on glycemic control and psychological aspects in persons with type 1 diabetes treated with sensor-augmented pump: A prospective single-center observational study.
Аннотация
AIMS/
INTRODUCTION
This study evaluated the effects of the Medtronic MiniMed 770G hybrid closed-loop system on glycemic control and psychological aspects in persons with type 1 diabetes mellitus.
MATERIALS AND
METHODS
This 3-month prospective observational study included 22 participants with type 1 diabetes mellitus who used the Medtronic MiniMed 640G predictive low-glucose suspend system and were switched to the 770G system. Time in the range of 70-180 mg/dL and glycated hemoglobin levels were evaluated; satisfaction, emotional distress and quality of life were assessed using self-reported questionnaires, including the Diabetes Treatment Satisfaction Questionnaire Status, Problem Area in Diabetes and Diabetes Therapy-Related Quality of Life.
RESULTS
Time in the range of 70-180 mg/dL increased (63.5 ± 13.4 to 73.0 ± 10.9% [mean ± standard deviation], P = 0.0010), and time above the range of 181-250 mg/dL decreased (26.9 ± 8.9 to 19.6 ± 7.1%, P < 0.0005). Glycated hemoglobin levels decreased (7.7 ± 1.0 to 7.2 ± 0.8%, P = 0.0021). The percentage of participants with time below the range of 54-69 mg/dL <4% of readings increased from 91% to 100% (P < 0.0005). No significant changes were detected in the satisfaction, emotional distress and quality of life levels, but increased sensor calibration might be related to worsened emotional distress and quality of life.
CONCLUSIONS
The hybrid closed-loop system decreased hyperglycemia and minimized hypoglycemia, but did not improve psychological aspects compared with the predictive low-glucose suspend system, probably because sensor calibration was increased.