DOI 10.3389/fneur.2025.1618737

Медицинская библиотека · Frontiers in neurology · 2025

Impact of stent retriever diameter on clinical outcomes in mechanical thrombectomy: a multicenter comparison of the 4-mm and 6-mm Solitaire devices.

Park Jung Hyun, Kim Hoon, Kim Sunghan

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

Optimal stent retriever diameter for mechanical thrombectomy in large vessel occlusion remains debated. We compared the efficacy and safety of Solitaire 4 × 40 mm and 6 × 40 mm stent retrievers in treating distal internal carotid artery (ICA) or middle cerebral artery (MCA) M1 segment occlusion.

METHODS

We retrospectively analyzed 145 patients with distal ICA or M1 occlusion treated with the Solitaire 4 × 40 mm (n = 59) or 6 × 40 mm (n = 86) devices. The primary outcome was the rate of first-pass modified thrombolysis in cerebral infarction (mTICI) ≥ 2b without rescue therapy. Secondary outcomes included the number of passes to achieve mTICI ≥2b, device switching and distal embolization rates, and safety parameters.

RESULTS

The 6 × 40 mm group displayed a significantly higher first-pass mTICI ≥2b rate (58.1% vs. 23.7%, p < 0.001), a lower number of passes to achieve successful revascularization (1.67 ± 1.00 vs. 2.54 ± 1.19, p < 0.001), and lower rates of device switching (4.7% vs. 18.6%, p = 0.015) and distal embolization (8.1% vs. 25.4%, p = 0.009). Postprocedural hemorrhage was less frequent in the 6 × 40 mm group (55.8% vs. 76.3%, p = 0.039), whereas symptomatic hemorrhage rates were similar between the groups (5.8% vs. 5.1%, p > 0.999).

CONCLUSION

The Solitaire 6 × 40 mm stent retriever showed more favorable efficacy outcomes and comparable safety compared to the 4 × 40 mm device for distal ICA or M1 occlusion, suggesting it may be considered a preferred option for these vessels.

Для цитирования: Park Jung Hyun, Kim Hoon, Kim Sunghan. Impact of stent retriever diameter on clinical outcomes in mechanical thrombectomy: a multicenter comparison of the 4-mm and 6-mm Solitaire devices.. Frontiers in neurology. 2025. DOI: 10.3389/fneur.2025.1618737.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.