DOI 10.1016/j.wneu.2019.04.234
Hemodynamic Fate of the Precommunicating Anterior Cerebral Artery Is Predicted by Vessel Dominance After Pipeline Embolization Device Deployment Across the Internal Carotid Artery Terminus.
Аннотация
BACKGROUND
The Pipeline embolization device (PED) is commonly used for the treatment of distal internal carotid artery (ICA) aneurysms, which often require placing the stent across the origin of the precommunicating segment of the anterior cerebral artery (ACA-A1). We sought to characterize the clinical and angiographic consequences of this maneuver.
METHODS
We performed a retrospective review of a prospectively collected database of patients treated with the PED at a single academic center from 2011 to 2017 to find patients for whom the PED was extended across the origin of the ACA-A1. The patient demographic data, pretreatment angiographic imaging findings, sizing of the bilateral A1 and ipsilateral M1 segment of the middle cerebral artery (MCA-M1), and follow-up angiographic and clinical imaging findings were recorded.
RESULTS
A total of 27 patients were included in the present study (8 men and 19 women; age, 52 ± 14.9 years). Follow-up angiography was conducted at a median of 9.2 months (interquartile range, 6; range, 5-84). The covered A1 segment was patent in 17 patients (63%). The covered ACA-A1/ipsilateral MCA-M1 ratio was 1.43 times greater for the patent ACA-A1 segments than those that were occluded (P = 0.0006). Similarly, the covered ACA-A1/contralateral ACA-A1 ratio was significantly larger statistically (1.64; P < 0.0001) for the patent ACA-A1 segments than that for those that were occluded. None of the patients developed clinical or radiographic signs of ACA stroke. The modified Rankin scale worsened for 1 patient during follow-up owing to a further decline of presenting vision loss.
CONCLUSIONS
The PED can be used to treat aneurysms with deployment from the MCA-M1 to the ICA without resulting in ACA stroke. Ipsilateral A1 segment dominance might be predictive of continued blood flow into the ACA after deployment at this location.