MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00701-023-05799-w

Медицинская библиотека · Acta neurochirurgica · 2023

Complications of deep brain stimulation in Parkinson's disease: a single-center experience of 517 consecutive cases.

Servello Domenico, Galbiati Tommaso Francesco, Iess Guglielmo, Minafra Brigida, Porta Mauro, Pacchetti Claudio

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

The number of deep brain stimulation (DBS) procedures is rapidly rising as well as the novel indications. Reporting adverse events related to surgery and to the hardware used is essential to define the risk-to-benefit ratio and develop novel strategies to improve it.

OBJECTIVE

To analyze DBS complications (both procedure-related and hardware-related) and further assess potential predictive factors.

METHODS

Five hundred seventeen cases of DBS for Parkinson's disease were performed between 2006 and 2021 in a single center (mean follow-up: 4.68 ± 2.86 years). Spearman's Rho coefficient was calculated to search for a correlation between the occurrence of intracerebral hemorrhage (ICH) and the number of recording tracks. Multiple logistic regression analyzed the probability of developing seizures and ICH given potential risk factors. Kaplan-Meier curves were performed to analyze the cumulative proportions of hardware-related complications.

RESULTS

Mortality rate was 0.2%, while permanent morbidity 0.6%. 2.5% of cases suffered from ICH which were not influenced by the number of tracks used for recordings. 3.3% reported seizures that were significantly affected by perielectrode brain edema and age. The rate of perielectrode brain edema was significantly higher for Medtronic's leads compared to Boston Scientific's (Χ2(1)= 5.927, P= 0.015). 12.2% of implants reported Hardware-related complications, the most common of which were wound revisions (7.2%). Internal pulse generator models with smaller profiles displayed more favorable hardware-related complication survival curves compared to larger designs (X2(1)= 8.139, P= 0.004).

CONCLUSION

Overall DBS has to be considered a safe procedure, but future research is needed to decrease the rate of hardware-related complications which may be related to both the surgical technique and to the specific hardware's design. The increased incidence of perielectrode brain edema associated with certain lead models may likewise deserve future investigation.

Для цитирования: Servello Domenico, Galbiati Tommaso Francesco, Iess Guglielmo, Minafra Brigida, Porta Mauro, Pacchetti Claudio. Complications of deep brain stimulation in Parkinson's disease: a single-center experience of 517 consecutive cases.. Acta neurochirurgica. 2023. DOI: 10.1007/s00701-023-05799-w.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.