DOI 10.1177/1352458517713667

Медицинская библиотека · Multiple sclerosis (Houndmills, Basingstoke, England) · 2017

Agreement analysis comparing iPad LCVA and Sloan testing in multiple sclerosis patients.

Sattarnezhad Neda, Farrow Samantha, Kimbrough Dorlan, Glanz Bonnie, Healy Brian, Chitnis Tanuja

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Visual symptoms are common in multiple sclerosis (MS). Low-contrast visual acuity (LCVA) testing using Sloan charts has demonstrated increased sensitivity for visual deficits compared to high-contrast acuity testing. Computerized testing of visual acuity may facilitate use in the clinic setting.

OBJECTIVES

To evaluate the agreement between an iPad-based and Sloan testing of LCVA in a cohort of MS patients.

METHODS

A total of 38 patients with relapsing-remitting MS were enrolled after providing informed written consent at Partners MS Center, Brigham and Women's hospital. Monocular LCVA was measured using retroilluminated Sloan chart and iPad-based LogMAR chart. Number of correct letters and agreement between two measurements were assessed for each eye using Bland-Altman analysis and paired t-test.

RESULTS

For both eyes, there was no significant difference in number correct between the two measurements using a paired t-test, and there was high correlation between two measurements (oculus dextrus (OD) r = 0.89, p < 0.001; oculus sinister (OS) r = 0.78, p < 0.001). The limits of agreement were -7.9 to +8.5 letters for the right eye and -10.9 to +11.2 letters for the left eye.

CONCLUSION

An iPad-based LCVA test shows good agreement with Sloan testing in MS patients.

Для цитирования: Sattarnezhad Neda, Farrow Samantha, Kimbrough Dorlan, Glanz Bonnie, Healy Brian, Chitnis Tanuja. Agreement analysis comparing iPad LCVA and Sloan testing in multiple sclerosis patients.. Multiple sclerosis (Houndmills, Basingstoke, England). 2017. DOI: 10.1177/1352458517713667.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.