MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00392-018-1288-1

Медицинская библиотека · Clinical research in cardiology : official journal of the German Cardiac Society · 2018

Acute and long-term outcomes of epicardial left atrial appendage ligation with the second-generation LARIAT device: a high-volume electrophysiology center experience.

Fink Thomas, Schlüter Michael, Tilz Roland Richard, Heeger Christian-Hendrik, Lemes Christine, Maurer Tilmann, Reissmann Bruno, Rottner Laura, Santoro Francesco, Mathew Shibu, Rillig Andreas, Ouyang Feifan, Kuck Karl-Heinz, Metzner Andreas

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Epicardial left atrial appendage (LAA) ligation may be an alternative to oral anticoagulation management and to endocardial LAA closure devices. We report long-term results after LAA ligation with the second-generation LARIAT device.

METHODS

Retrospective study was performed on patients who underwent LAA ligation at our center. Follow-up included patient visits and transesophageal echocardiography (TEE) to assess LAA-to-LA leakages.

RESULTS

76 patients with an indication for LAA closure underwent cardiac CT-based screening with 16 patients (21%) excluded from LAA ligation due to anatomical reasons. Finally, 48 patients (70 ± 9 years, 23 women) underwent LAA ligation with successful LAA closure in 44 patients (92%). Major periprocedural complications occurred in three patients (6%; one major femoral bleeding, one pneumothorax with surgical drainage, one right ventricular perforation with concomitant stroke). Additionally, minor complications occurred in 27% of the patients with mild pericarditis in eight of these patients being the most common adverse event. Clinical follow-up (median of 443 days, interquartile range 158; 773) was obtained from 38 patients. One thromboembolic event (transient ischemic attack) occurred. Complete LAA closure was demonstrated in 23/35 patients (66%) with TEE follow-up. Major leakages > 5 mm without documentation of intracardiac thrombi were documented in four patients (11%).

CONCLUSIONS

Epicardial LAA ligation with the second-generation LARIAT device was associated with a high acute success rate comparable to endocardial LAA closure devices. LAA ligation was accompanied by a relevant incidence of periprocedural complications with mild pericarditis being the most common adverse event. Follow-up demonstrated a moderate incidence of leakages after LAA exclusion, no intracardiac thrombus formation and only a single thromboembolic event. The effectiveness in preventing thromboembolic events needs further investigation in larger patient cohorts.

Для цитирования: Fink Thomas, Schlüter Michael, Tilz Roland Richard, Heeger Christian-Hendrik, Lemes Christine, Maurer Tilmann, Reissmann Bruno, Rottner Laura, Santoro Francesco, Mathew Shibu, Rillig Andreas, Ouyang Feifan, Kuck Karl-Heinz, Metzner Andreas. Acute and long-term outcomes of epicardial left atrial appendage ligation with the second-generation LARIAT device: a high-volume electrophysiology center experience.. Clinical research in cardiology : official journal of the German Cardiac Society. 2018. DOI: 10.1007/s00392-018-1288-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.