Bioabsorbable Drug-Eluting Stent Versus Bare Metal Stent in Iliac Artery Evaluated by Optical Coherence Tomography: An In Vivo Study in Porcine.
Аннотация
OBJECTIVE
This study aimed to compare, using optical coherence tomography (OCT), the outcomes of bioabsorbable drug-eluting stent with those of bare metal stent (BMS) following implantation in porcine iliac artery.
METHODS
After the placement of BMS and bioabsorbable drug-eluting stents, we used OCT and digital subtraction angiography to investigate stent appositions, arterial neointima, evagination, and restenosis at 1 and 3 months.
RESULTS
At 1 and 3 months after stent implantation, OCT study was performed to investigate 32 stents and 21 788 struts. Thirty-three malapposed struts were found in the bioabsorbable drug-eluting stent groups and 2 were found in BMS groups. The average neointimal thickness, area, and in-stent stenosis were significantly lower in bioabsorbable drug-eluting stents than in BMS, while the frequency of malapposed struts was higher in the bioresorbable drug-eluting stent groups. Average neointimal thickness was lower in bioabsorbable drug-eluting stents than in BMS at 1 (0.19 ± 0.09 vs 0.67 ± 0.75 mm; P < .001) and 3 months (0.21 ± 0.08 vs 1.52 ± 0.28 mm; P < .001).
CONCLUSION
Our study suggested that bioabsorbable drug-eluting stent is more effective in decreasing arterial restenosis than BMS in animal models.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.