MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/j.1755-3768.2009.01605.x

Медицинская библиотека · Acta ophthalmologica · 2011

Complications of intrastromal corneal ring segment implantation using a femtosecond laser for channel creation: a survey of 850 eyes with keratoconus.

Coskunseven Efekan, Kymionis George D, Tsiklis Nikolaos S, Atun Serife, Arslan Ebru, Siganos Charalambos S, Jankov Mirco, Pallikaris Ioannis G

Клиническое исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To report complications after the implantation of intrastromal corneal ring segments (Keraring; Mediphacos, Belo Horizonte, Brazil) in keratoconic eyes using a femtosecond laser (IntraLase; Advanced Medical Optics, Santa Ana, California, USA) for channel creation.

METHODS

Retrospective chart review of 531 patients (850 eyes) who underwent Keraring insertion using a femtosecond laser for channel creation. Intraoperative and postoperative complications were recorded.

RESULTS

Intraoperatively, there were 22 (2.7%) cases of incomplete channel formation. The rest of the intraoperative complications were galvo lag error (system malfunction) [five eyes (0.6%)], endothelial perforation [five eyes (0.6%)] and incorrect entry of the channel [two eyes (0.2%)]. Postoperatively, there were 11 (1.3%) cases of segment migration, two (0.2%) cases of corneal melting and one (0.1%) case of mild infection. The overall complication rate was 5.7% (49 cases out of 850 eyes).

CONCLUSION

In this study, intracorneal ring segment implantation using a femtosecond laser for channel creation was related to a number of complications. The most common complications were incomplete channel creation (intraoperatively) and segment migration (postoperatively).

Для цитирования: Coskunseven Efekan, Kymionis George D, Tsiklis Nikolaos S, Atun Serife, Arslan Ebru, Siganos Charalambos S, Jankov Mirco, Pallikaris Ioannis G. Complications of intrastromal corneal ring segment implantation using a femtosecond laser for channel creation: a survey of 850 eyes with keratoconus.. Acta ophthalmologica. 2011. DOI: 10.1111/j.1755-3768.2009.01605.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.