MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3233/thc-182509

Медицинская библиотека · Technology and health care : official journal of the European Society for Engineering and Medicine · 2018

Experimental investigation of the determination for the safe operating regime of ultrasound tube-shaped waveguide wire for internal blood vessel debulking.

Bubulis Algimantas, Venslauskas Mantas, Navickas Jonas, Abramavičius Silvijus, Stankevičius Edgaras

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Majority of limb amputations are caused by circulatory disturbances such as vascular occlusions and strictures. Discovery of modern and more advanced ultrasonic interventional vascular debulking methodology would likely save limbs of CVD patients and their lives in an economical way. However, there is a lack of researches regarding the ultrasound's effect on physiological functions of human blood cells. The tube-shaped ultrasound waveguide wire with orifices at its operational end was offered as the alternative to some currently patented interventional thrombosis treatment solutions.

OBJECTIVE

To establish the safe operating regime of the proposed device.

METHODS

The temperature rise induced by the cavitation process and friction between the waveguide and surrounding fluids was measured and microscopic pictures of human blood were made.

RESULTS

Blood insonation lasting 15 seconds, leads to blood clot formation. If insonation continues for 30 seconds some cells are totally destroyed. In addition, the safe operating regime was established. To avoid heating of the environment to the temperature harmful for the medium (blood) and surrounding tissues, is achieved when the system should be on for 40%, and of for 60% of the period of 1 second.

CONCLUSIONS

The safe operating regime of the proposed device was established.

Для цитирования: Bubulis Algimantas, Venslauskas Mantas, Navickas Jonas, Abramavičius Silvijus, Stankevičius Edgaras. Experimental investigation of the determination for the safe operating regime of ultrasound tube-shaped waveguide wire for internal blood vessel debulking.. Technology and health care : official journal of the European Society for Engineering and Medicine. 2018. DOI: 10.3233/thc-182509.