Measurement of depth of Intacs implanted via femtosecond laser using Pentacam.

Kamburoglu Günhal, Ertan Aylin, Saraçbasi Osman

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To measure the depth of intracorneal ring segments (Intacs, Addition Technology Inc) implanted via femtosecond laser by Pentacam rotating Scheimpflug camera (Oculus Optikgeräte GmbH) and to determine the change in depth at 1-year follow-up.

METHODS

Thirty-two eyes of 23 keratoconic patients receiving Intacs intrastromal corneal ring segments implanted by femtosecond laser were included in the study. The depth of intracorneal channel and Intacs segments were measured for 7 predetermined points at 1 week and 1 year postoperatively using Scheimpflug images.

RESULTS

Mean achieved intrastromal channel depth exceeded the intended parameter by 6.5 microm to 69.0 microm. The standard deviation varied between +/- 13.5 microm and +/- 44.2 microm. The segment depth decreased at the end of the first postoperative year in all measured points, and the change was statistically significant at the superior, inferior, and temporal sides of the Intacs. The degree of change was not correlated with size of the Intacs, preoperative central corneal thickness, and mean keratometric values.

CONCLUSIONS

Pentacam provided imaging of the Intacs segments and measurement of their depth, enabling evaluation of the changes in depth at 1-year follow-up. The Intacs segments tended to shallow over time, which may be a factor for depth-related complications.

Для цитирования: Kamburoglu Günhal, Ertan Aylin, Saraçbasi Osman. Measurement of depth of Intacs implanted via femtosecond laser using Pentacam.. Journal of refractive surgery (Thorofare, N.J. : 1995). 2009. DOI: 10.3928/1081597x-20090401-05.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.