MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/ijg.0b013e3181622502

Медицинская библиотека · Journal of glaucoma · 2008

Assessment of the pascal dynamic contour tonometer in measuring intraocular pressure in keratoconic eyes.

Papastergiou Georgios I, Kozobolis Vasiliios, Siganos Dimitrios S

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To assess the performance of the Pascal dynamic contour tonometer (PDCT) (Swiss Microtechnology AG) by comparing it to Goldmann applanation tonometry (GAT) and noncontact air tonometry (NCT) in keratoconic eyes.

METHODS

Intraocular pressure (IOP) was measured in both eyes of 32 patients with keratoconus using PDCT, GAT, and NCT. IOP measurements were also obtained with all 3 instruments in 46 patients with healthy eyes in the same age range, which served as controls. Keratometric measurements were obtained in all eyes using Orbscan II (Bausch & Lomb) and central corneal thickness was measured using Corneo-gaze plus (Sonogaze).

RESULTS

In the control group, a correlation between PDCT, GAT, and NCT IOP measurements was observed. In keratoconic eyes, GAT and NCT IOP measurements were significantly lower than PDCT measurements (5.3+/-2.2 mm Hg and 4.75+/-1.7 mm Hg, respectively). Unlike GAT and NCT, PDCT measurements in both groups were not significantly affected by central corneal thickness. Additionally, the mean IOP measurement in the keratoconic eyes was lower than that of the control group, particularly when measured with GAT and NCT. Corneal curvature, did not significantly affect IOP measurements of any instrument in either group.

CONCLUSIONS

The corneal abnormalities characteristical of keratoconus seem to influence more GAT and NCT IOP measurements than PDCT IOP measurements. Therefore, PDCT may be better suited for monitoring IOP in keratoconic eyes.

Для цитирования: Papastergiou Georgios I, Kozobolis Vasiliios, Siganos Dimitrios S. Assessment of the pascal dynamic contour tonometer in measuring intraocular pressure in keratoconic eyes.. Journal of glaucoma. 2008. DOI: 10.1097/ijg.0b013e3181622502.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.