Accuracy of Goldmann, ocular response analyser, Pascal and TonoPen XL tonometry in keratoconic and normal eyes.

Mollan S P, Wolffsohn J S, Nessim M, Laiquzzaman M, Sivakumar S, Hartley S, Shah S

Сравнительное исследованиеОценочное исследованиеСтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

AIM: The aim of this study was to evaluate the practicality and accuracy of tonometers used in routine clinical practice for established keratoconus (KC).

METHODS

This was a prospective study of 118 normal and 76 keratoconic eyes where intraocular pressure (IOP) was measured in random order using the Goldman applanation tonometer (GAT), Pascal dynamic contour tonometer (DCT), Reichert ocular response analyser (ORA) and TonoPen XL tonometer. Corneal hysteresis (CH) and corneal resistance factor (CRF), as calculated by the ORA, were recorded. Central corneal thickness (CCT) was measured using an ultrasound pachymeter.

RESULTS

The difference in IOP values between instruments was highly significant in both study groups (p<0.001). All other IOP measures were significantly higher than those for GAT, except for the Goldmann-correlated IOP (average of the two applanation pressure points) (IOPg) as measured by ORA in the control group and the CH-corrected IOP (corneal-compensated IOP value) (IOPcc) measures in the KC group. CCT, CH and CRF were significantly less in the KC group (p<0.001). Apart from the DCT, all techniques tended to measure IOP higher in eyes with thicker corneas.

CONCLUSION

The DCT and the ORA are currently the most appropriate tonometers to use in KC for the measurement of IOPcc. Corneal factors such as CH and CRT may be of more importance than CCT in causing inaccuracies in applanation tonometry techniques.

Для цитирования: Mollan S P, Wolffsohn J S, Nessim M, Laiquzzaman M, Sivakumar S, Hartley S, Shah S. Accuracy of Goldmann, ocular response analyser, Pascal and TonoPen XL tonometry in keratoconic and normal eyes.. The British journal of ophthalmology. 2008. DOI: 10.1136/bjo.2007.136473.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.