m+Medelevator.RU
← Поиск публикаций
Journal of cataract and refractive surgery · 2008

Repeatability of corneal first-surface wavefront aberrations measured with Pentacam corneal topography.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Shankar Hema, Taranath Deepa, Santhirathelagan Chandramalar T, Pesudovs Konrad

Аннотация

PURPOSE: To assess the repeatability of corneal wavefront aberrations derived from Pentacam (Oculus) corneal topography.

SETTING: Flinders Eye Centre, Flinders Medical Centre, Bedford Park, South Australia, Australia.

METHODS: Forty-five normal participants and 10 participants with keratoconus were tested. Intraobserver and interobserver repeatability was determined using 4 observers within and between sessions. Topographical maps were exported to external software, and corneal first-surface wavefront aberrations were calculated using a 10th-order Zernike expansion over a 6.0 mm optical zone. Repeatability was determined with Bland-Altman limits of agreement and expressed as the coefficient of repeatability (COR).

RESULTS: Initial data showed high wavefront aberrations in normal participants and poor repeatability. Topographical maps showed extrapolated topography in zones without data acquisition; maps with less than 6.0 mm of complete data were excluded in the final analysis. The mean wavefront aberrations for normal participants remained high, but repeatability improved. The COR relative to the magnitude of wavefront aberrations was high (average 100%) across all modal pairs and orders, although best for total higher-order root mean square. Participants with keratoconus had higher magnitude wavefront aberrations and poorer repeatability but similar COR to average wavefront aberration ratios. Examination of raw elevation data showed poor repeatability.

CONCLUSIONS: Wavefront aberrations calculated from Pentacam corneal topography were large in magnitude, and reliability was poor, largely due to variability in corneal elevation data. Intraobserver and interobserver reliability within and between sessions was comparable. The Pentacam was not reliable in measuring corneal wavefront aberrations.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты