DOI 10.1111/j.1540-8159.1988.tb06298.x
The combined use of hysteresis and Holter functions improves diagnosis and therapy in patients with sick sinus syndrome.
Аннотация
We used the Holter functions of an implantable pacemaker in 12 patients with sino-atrial disease suffering from Adam-Stokes attacks to analyze the effects of hysteresis programming. The basic pacing rate of 50 ppm combined with a programmed hysteresis value of 10 to 20 ppm led to a drastic reduction in competition between paced and spontaneous beats: the percentage of paced beats decreased from 20.1% without hysteresis to 1.4% using 10 ppm hysteresis and the switches from sense to pace decreased from 2.0% to 0.06%. On the other hand, with hysteresis, no decrease in either the exercise tolerance or general well-being of the patients was observed.