MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/j.1540-8159.1988.tb06293.x

Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 1988

Noninvasive measurement of retrograde conduction times in pacemaker patients.

Marco D D, Gallagher D

СтатьяПеревод пока не подготовлен

Аннотация

Retrograde, or ventriculoatrial conduction (VAC) has been measured in 67% of pacemaker patients with SSS and 25% of patients with 3 degrees block. DDD pacers use the postventricular atrial refractory period (PVARP) to prevent sensing VAC. If the PVARP is too short, a pacer-mediated tachycardia (PMT) could occur. If too long, rapid atrial rates may not be sensed. To accurately set the PVARP, VAC must be measured. Five noninvasive methods are described: (1) Surface ECG during VVI pacing; (2) Atrial intracardiac electrograms during VVI pacing; (3) Initiating a PMT with long pacemaker AV interval; (4) Initiating a PMT at the maximum tracking rate; (5) DDI mode with electronic calipers. The description of these methods is intended to enable a physician to detect and measure VAC in patients with any manufacturer's DDD or VDD pacemaker.

Для цитирования: Marco D D, Gallagher D. Noninvasive measurement of retrograde conduction times in pacemaker patients.. Pacing and clinical electrophysiology : PACE. 1988. DOI: 10.1111/j.1540-8159.1988.tb06293.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.