Comparison of different correction formulas and measurement methods for the accurate determination of intraocular pressure after SMILE and FS-LASIK surgery.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Yang Zhiqing, Miao Na, Wang Lixiang, Ma Ke
Аннотация
BACKGROUND: To compare the accuracy of non-contact tonometry, Pentacam, and corneal visualization Scheimpflug technology (Corvis ST) for the measurement of intraocular pressure (IOP) after small incision lenticule extraction (SMILE) or femtosecond laser-assisted in situ keratomileusis (FS-LASIK) surgery.
METHODS: A total of 49 patients (98 eyes) undergoing FS-LASIK or SMILE surgery at West China Hospital, Sichuan University from January to March 2021 were enrolled in this prospective, comparative, self-controlled study. IOP values were measured with non-contact tonometer, Pentacam, and Corvis ST before surgery and 1 month after surgery. Pentacam-derived postoperative IOP values were corrected using five correction formulas (Ehlers, Shah, Dresden, Orssengo-Pye, and Kohlhaas), while Corvis ST-derived values were corrected using a single formula.
RESULTS: In the SMILE group, no significant differences were observed between the IOP values corrected with the Shah formula before and after surgery (t = 0.17, P = 0.869), whereas postoperative IOP values corrected with the other formulas were significantly different from the corresponding preoperative measurements (P < 0.05). In the FS-LASIK group, postoperative IOP values corrected with the Ehlers, Shah, or Corvis ST formulas were significantly different from the corresponding preoperative IOP measurements (P < 0.05), but no significant differences were observed between pre- and postoperative IOP values corrected with the Dresden (t = - 0.08, P = 0.941), Orssengo-Pye (t = - 0.52, P = 0.604), or Kohlhaas (t = 1.22, P = 0.231) formulas.
CONCLUSIONS: Pentacam's Shah correction formula seemed to be the most appropriate method for accurately measuring postoperative IOP in patients undergoing SMILE surgery, while the Dresden, Orssengo-Pye, and Kohlhaas correction formulas of Pentacam were identified as the most reliable methods for estimating IOP in patients after FS-LASIK surgery.
METHODS: A total of 49 patients (98 eyes) undergoing FS-LASIK or SMILE surgery at West China Hospital, Sichuan University from January to March 2021 were enrolled in this prospective, comparative, self-controlled study. IOP values were measured with non-contact tonometer, Pentacam, and Corvis ST before surgery and 1 month after surgery. Pentacam-derived postoperative IOP values were corrected using five correction formulas (Ehlers, Shah, Dresden, Orssengo-Pye, and Kohlhaas), while Corvis ST-derived values were corrected using a single formula.
RESULTS: In the SMILE group, no significant differences were observed between the IOP values corrected with the Shah formula before and after surgery (t = 0.17, P = 0.869), whereas postoperative IOP values corrected with the other formulas were significantly different from the corresponding preoperative measurements (P < 0.05). In the FS-LASIK group, postoperative IOP values corrected with the Ehlers, Shah, or Corvis ST formulas were significantly different from the corresponding preoperative IOP measurements (P < 0.05), but no significant differences were observed between pre- and postoperative IOP values corrected with the Dresden (t = - 0.08, P = 0.941), Orssengo-Pye (t = - 0.52, P = 0.604), or Kohlhaas (t = 1.22, P = 0.231) formulas.
CONCLUSIONS: Pentacam's Shah correction formula seemed to be the most appropriate method for accurately measuring postoperative IOP in patients undergoing SMILE surgery, while the Dresden, Orssengo-Pye, and Kohlhaas correction formulas of Pentacam were identified as the most reliable methods for estimating IOP in patients after FS-LASIK surgery.