Which keratometer is most reliable for correcting astigmatism with toric intraocular lenses?
Аннотация
PURPOSE: To evaluate the accuracy of preoperative keratometers used in cataract surgery with toric intraocular lens (IOL).
METHODS
Twenty-five eyes received an AcrySof toric IOL implantation. Four different keratometric methods, a manual keratometer, an IOL master, a Pentacam and an auto keratometer, were performed preoperatively in order to evaluate preexisting corneal astigmatism. Differences between the true residual astigmatism and the anticipated residual astigmatism (keratometric error) were compared at one and three months after surgery by using a separate vector analysis to identify the keratometric method that provided the highest accuracy for astigmatism control.
RESULTS
The mean keratomeric error was 0.52 diopters (0.17-1.17) for the manual keratometer, 0.62 (0-1.31) for the IOL master, 0.69 (0.08-1.92) for the Pentacam, and 0.59 (0.08-0.94) for the auto keratometer. The manual keratometer was the most accurate, although there was no significant difference between the keratometers (p > 0.05). All of the keratometers achieved an average keratometric error of less than one diopter.
CONCLUSIONS
Manual keratometry was the most accurate of the four methods evaluated, although the other techniques were equally satisfactory in determining corneal astigmatism.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.