Simulation of toric intraocular lens results: manual keratometry versus dual-zone automated keratometry from an integrated biometer.

Hill Warren, Osher Robert, Cooke David, Solomon Kerry, Sandoval Helga, Salas-Cervantes Rosario, Potvin Richard

Сравнительное исследованиеСтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate simulated clinical outcomes in patients with toric intraocular lenses (IOLs) calculated on the basis of dual-zone automated keratometry from an integrated optical biometer, relative to manual keratometry.

SETTING

Private practice, Mesa, Arizona, USA.

DESIGN

Comparative case series.

METHODS

Patient records at 4 clinical sites were reviewed to identify patients who had manual keratometry and biometry with the Lenstar LS 900 recorded before toric IOL implantation and refractive follow-up data after implantation. Preoperative and operative data were extracted from patient charts. Simulated refractive outcomes were calculated based on mathematically removing the actual IOL implanted and then mathematically inserting the IOLs as determined by manual or automated keratometry from the biometry device.

RESULTS

Data for 128 patients were available for analysis. The actual residual astigmatism was comparable between manual keratometry and automated keratometry from the biometry system. Although simulated residual refractive astigmatism was similar between the 2 devices on average, there was variability in results by patient. Simulated residual refractive astigmatism was lower for the biometer when the standard deviation of the angle of astigmatism was low. Site-to-site variability was lower with the biometer than with manual keratometry.

CONCLUSIONS

Simulated outcomes suggest that overall results for a group of patients whose toric IOL surgery planning is performed with the dual-zone automated keratometry data from the biometer will be equivalent to those when manual keratometry is used. The reduced site-to-site variability with the biometer suggests an operational advantage.

Для цитирования: Hill Warren, Osher Robert, Cooke David, Solomon Kerry, Sandoval Helga, Salas-Cervantes Rosario, Potvin Richard. Simulation of toric intraocular lens results: manual keratometry versus dual-zone automated keratometry from an integrated biometer.. Journal of cataract and refractive surgery. 2011. DOI: 10.1016/j.jcrs.2011.06.028.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.