DOI 10.1016/j.medengphy.2025.104408
Biomechanical evaluation of intelligent fluid-solid coupling vaginal dilatation system: Experimental and numerical analysis.
Аннотация
OBJECTIVES
Cervical cancer is a serious threat to women's life and health and has a high mortality rate. Colposcopy is an important method for early clinical cervical cancer screening, but the traditional vaginal dilator has problems such as discomfort in use and cumbersome operation. For this reason, this study aims to design an intelligent vaginal dilatation system to automate colposcopy and enhance patient comfort.
METHODS
An intelligent vaginal dilatation system combining flexible and rigid dilatation techniques is proposed. A fluid-solid coupled finite element method was used to simulate the dilation process of the system during colposcopy. The smart dilator was inserted into a simulated vaginal model in the simulation, and the fluid domain pressure inlet was set to be 1.5 kPa, and the solid material was a hyperelastic model of medical silicone rubber. Subsequently, the prototype machining was completed and in vitro dilatation displacement and pressure experiments were conducted.
RESULTS
The simulation results showed that the maximum expansion of the system was 32.2 mm, and the average pressure on the simulated vaginal wall was 605.91 kPa. The average maximum expansion of the system in the in-vitro expansion displacement experiment was 30.49 ± 0.05 mm, which was basically the same as the simulation results. The results of the in vitro pressure experiment showed that the intelligent dilatation system had a larger force area on the vaginal wall at the same level of dilatation, and the pressure value was smaller and more uniformly distributed. Compared with the traditional duckbill dilator, it can effectively reduce the local pressure feeling and improve the uniformity of dilation.
CONCLUSION
The intelligent vaginal dilatation system proposed in this study is superior to traditional dilators in terms of dilatation performance, safety and comfort. The feasibility of its design and potential for clinical application were verified.