Peripheral visual field shifts after intraocular lens implantation.

van Vught Luc, Luyten Gregorius P M, Beenakker Jan-Willem M

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To assess whether intraocular lens (IOL) implantation induces shifts in the peripheral visual field.

SETTING

Department of Ophthalmology, Leiden University Medical Center, Leiden, the Netherlands.

DESIGN

Ray-tracing study.

METHODS

Nonsequential ray-tracing simulations were performed with phakic and pseudophakic versions of the same eye model to assess potential shifts in the visual field after IOL implantation. 2 different IOL designs were evaluated and for each design 5 different axial positions and 7 different intrinsic powers were tested. The relation between the physical position of the light source and the location where the retina was illuminated was determined for each eye model. Subsequently, these relations were used to calculate whether the visual field shifts in pseudophakic eyes.

RESULTS

The pseudophakic visual field shift was below 1 degree for central vision in all evaluated models. For peripheral vision, the light rays in the pseudophakic eyes were refracted to a more central retinal location compared with phakic eyes, resulting in a central shift of the peripheral visual field. The magnitude of the shift depended on the IOL design and its axial position, but could be as high as 5.4 degrees towards central vision.

CONCLUSIONS

IOL implantation tends to have little effect on the central visual field but can induce an over 5 degrees shift in the peripheral visual field. Such a shift can affect the perception of peripheral visual complaints.

Для цитирования: van Vught Luc, Luyten Gregorius P M, Beenakker Jan-Willem M. Peripheral visual field shifts after intraocular lens implantation.. Journal of cataract and refractive surgery. 2023. DOI: 10.1097/j.jcrs.0000000000001299.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.