MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2147/opth.s417357

Медицинская библиотека · Clinical ophthalmology (Auckland, N.Z.) · 2023

Intraocular Lens Unfurling Time Exponentially Decays with Increased Solution Temperature.

Rocher Erick E, Mukherjee Rishima, Pitingolo James, Levenshus Eli, Alexander Gwyneth, Park Minyoung, Acharya Rupsa, Khan Sarah, Shuff Jordan, Aguirre Andres, Matin Shababa, Walter Keith, Eghrari Allen O

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Intraocular lens (IOL) unfurling can be a rate-limiting step in cataract surgery, limiting operative efficiency. Furthermore, inefficient unfurling has important implications for clinical outcomes. We examine the effects of solution temperature on IOL unfurling time using three in vitro models of the ocular environment.

METHODS

IOLs were injected into a 6-well plate filled with balanced salt solution (BSS), dispersive ophthalmic viscoelastic device (OVD), or cohesive OVD. Experiments were also performed in a plastic eye filled with dispersive or cohesive OVD. IOL unfurling time was recorded against the temperature of the respective solution.

RESULTS

IOL unfurling time decayed exponentially as solution temperature increased in all experiments, including the BSS-filled 6-well plate, the OVD-filled 6-well plate, and the OVD-filled plastic eye. IOLs failed to unfurl within 10 min at 10°C, below the glass transition temperature of the tested IOLs. Increasing solution temperature from 20°C to 30°C decreases IOL unfurling by greater than 2 min. Further heating to 40°C did not significantly decrease IOL unfurling time.

CONCLUSION

Increased solution temperature rapidly decreases IOL unfurling time in vitro. IOLs do not unfurl within a clinically acceptable timeframe at or below their glass transition temperature. Increased BSS and/or OVD temperature may be a potential method to decrease IOL unfurling time in cataract surgery. However, future research is needed to elucidate potential consequences of warmed BSS and/or OVD on post-operative outcomes. This study demonstrates the potential for temperature regulation to decrease cataract surgery operative time and provides preliminary evidence to justify future clinical validation of this relationship.

Для цитирования: Rocher Erick E, Mukherjee Rishima, Pitingolo James, Levenshus Eli, Alexander Gwyneth, Park Minyoung, Acharya Rupsa, Khan Sarah, Shuff Jordan, Aguirre Andres, Matin Shababa, Walter Keith, Eghrari Allen O. Intraocular Lens Unfurling Time Exponentially Decays with Increased Solution Temperature.. Clinical ophthalmology (Auckland, N.Z.). 2023. DOI: 10.2147/opth.s417357.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.