DOI 10.1093/ejo/cju103

Медицинская библиотека · European journal of orthodontics · 2015

Impact of orthodontic appliances on the quality of craniofacial anatomical magnetic resonance imaging and real-time speech imaging.

Wylezinska Marzena, Pinkstone Marie, Hay Norman, Scott Andrew D, Birch Malcolm J, Miquel Marc E

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

AIMS: The aim of this work was to investigate the effects of commonly used orthodontic appliances on the magnetic resonance (MR) image quality of the craniofacial region, with special interest in the soft palate and velopharyngeal wall using real-time speech imaging sequences and anatomical imaging of the temporomandibular joints (TMJ) and pituitaries.

METHODS

Common orthodontic appliances were studied on 1.5 T scanner using standard spin and gradient echo sequences (based on the American Society for Testing and Materials standard test method) and sequences previously applied for high-resolution anatomical and dynamic real-time imaging during speech. Images were evaluated for the presence and size of artefacts.

RESULTS

Metallic orthodontic appliances had different effects on image quality. The most extensive individual effects were associated with the presence of stainless steel archwire, particularly if combined with stainless steel brackets and stainless steel molar bands. With those appliances, diagnostic quality of magnetic resonance imaging speech and palate images will be most likely severely degraded, or speech imaging and imaging of pituitaries and TMJ will be not possible. All non-metallic, non-metallic with Ni/Cr reinforcement or Ni/Ti alloys appliances were of little concern.

LIMITATIONS: The results in the study are only valid at 1.5 T and for the sequences and devices used and cannot necessarily be extrapolated to all sequences and devices. Furthermore, both geometry and size of some appliances are subject dependent, and consequently, the effects on the image quality can vary between subjects. Therefore, the results presented in this article should be treated as a guide when assessing the risks of image quality degradation rather than an absolute evaluation of possible artefacts.

CONCLUSIONS

Appliances manufactured from stainless steel cause extensive artefacts, which may render image non-diagnostic. The presence and type of orthodontic appliances should be always included in the patient's screening, so the risks of artefacts can be assessed prior to imaging. Although the risks to patients with fixed orthodontic appliances at 1.5 T MR scanners are low, their secure attachment should be confirmed prior to the examination.

Для цитирования: Wylezinska Marzena, Pinkstone Marie, Hay Norman, Scott Andrew D, Birch Malcolm J, Miquel Marc E. Impact of orthodontic appliances on the quality of craniofacial anatomical magnetic resonance imaging and real-time speech imaging.. European journal of orthodontics. 2015. DOI: 10.1093/ejo/cju103.