A preclinical Talbot-Lau prototype for x-ray dark-field imaging of human-sized objects.

Hauke C, Bartl P, Leghissa M, Ritschl L, Sutter S M, Weber T, Zeidler J, Freudenberger J, Mertelmeier T, Radicke M, Michel T, Anton G, Meinel F G, Baehr A, Auweter S, Bondesson D, Gaass T, Dinkel J, Reiser M, Hellbach K

СтатьяВалидационное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: Talbot-Lau x-ray interferometry provides information about the scattering and refractive properties of an object - in addition to the object's attenuation features. Until recently, this method was ineligible for imaging human-sized objects as it is challenging to adapt Talbot-Lau interferometers (TLIs) to the relevant x-ray energy ranges. In this work, we present a preclinical Talbot-Lau prototype capable of imaging human-sized objects with proper image quality at clinically acceptable dose levels.

METHODS

The TLI is designed to match a setup of clinical relevance as closely as possible. The system provides a scan range of 120 × 30 cm2 by using a scanning beam geometry. Its ultimate load is 100 kg. High aspect ratios and fine grid periods of the gratings ensure a reasonable setup length and clinically relevant image quality. The system is installed in a university hospital and is, therefore, exposed to the external influences of a clinical environment. To demonstrate the system's capabilities, a full-body scan of a euthanized pig was performed. In addition, freshly excised porcine lungs with an extrinsically provoked pneumothorax were mounted into a human thorax phantom and examined with the prototype.

RESULTS

Both examination sequences resulted in clinically relevant image quality - even in the case of a skin entrance air kerma of only 0.3 mGy which is in the range of human thoracic imaging. The presented case of a pneumothorax and a reader study showed that the prototype's dark-field images provide added value for pulmonary diagnosis.

CONCLUSION

We demonstrated that a dedicated design of a Talbot-Lau interferometer can be applied to medical imaging by constructing a preclinical Talbot-Lau prototype. We experienced that the system is feasible for imaging human-sized objects and the phase-stepping approach is suitable for clinical practice. Hence, we conclude that Talbot-Lau x-ray imaging has potential for clinical use and enhances the diagnostic power of medical x-ray imaging.

Для цитирования: Hauke C, Bartl P, Leghissa M, Ritschl L, Sutter S M, Weber T, Zeidler J, Freudenberger J, Mertelmeier T, Radicke M, Michel T, Anton G, Meinel F G, Baehr A, Auweter S, Bondesson D, Gaass T, Dinkel J, Reiser M, Hellbach K. A preclinical Talbot-Lau prototype for x-ray dark-field imaging of human-sized objects.. Medical physics. 2018. DOI: 10.1002/mp.12889.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.