MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1378/chest.115.5.1333

Медицинская библиотека · Chest · 1999

Accuracy of respiratory inductive plethysmography for the diagnosis of upper airway resistance syndrome.

Loube D I, Andrada T, Howard R S

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To determine the sensitivity and specificity of quantitative respiratory inductive plethysmography (RIP) compared with the "gold standard," nocturnal esophageal pressure (Pes) measurement, in the diagnosis of upper airway resistance syndrome (UARS) in adults.

METHODS

Fourteen consecutive patients without obstructive sleep apnea and suspected of having UARS underwent simultaneous measurement of Pes with a catheter and standard nocturnal polysomnography along with RIP. UARS events (RERAs, respiratory effort-related arousals) were identified by observing crescendo changes in Pes with a Pes nadir < or = -12 cm H2O, followed by an arousal or microarousal. UARS was defined as > or = 10 RERAs per hour. For each patient, the ratio of peak inspiratory flow to mean inspiratory flow (PIFMF) measured by RIP was performed during quiet wakefulness and with 40 randomly selected breaths in the supine position for two conditions: stage 2 sleep, immediately prior to arousals in any sleep stage. The mean PIFMF (wake-sleep) was calculated for each condition.

RESULTS

The sensitivities and specificities, respectively, of RIP to distinguish UARS patients from non-UARS patients are from stage 2 sleep (67%, 80%), immediately prior to arousals (100%, 100%). For breaths occurring immediately prior to arousals, the mean PIFMF (wake-sleep) is > or = 0.13 for UARS patients and < 0.13 for non-UARS patients.

CONCLUSION

The PIFMF measured by RIP allows for the most accurate identification of UARS patients when breaths are selected for analysis immediately prior to arousals.

Для цитирования: Loube D I, Andrada T, Howard R S. Accuracy of respiratory inductive plethysmography for the diagnosis of upper airway resistance syndrome.. Chest. 1999. DOI: 10.1378/chest.115.5.1333.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.