Quantitative validation of data-driven motion correction for brain PET using phantom with motion generator system.

Kamitaka Yuto, Sakata Muneyuki, Oda Keiichi, Katsuki Akie, Kawakami Hirofumi, Wagatsuma Kei, Kobayashi Masato, Ishii Kenji

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Head motion during brain positron emission tomography (PET) degrades image quality and quantitative accuracy. Therefore, a data-driven motion correction (MC) method utilizing ultrafast list-mode reconstruction technology has been proposed and shown to considerably improve image quality. However, reproducing accurate actual motions and motion-free images using clinical data alone remains challenging. This study aimed to quantitatively evaluate data-driven MC using a brain phantom for known tracer distributions and a custom-made motion generator system for variable known motions.

METHODS

Hoffman 3D brain phantom was filled with 20 and 3 MBq of [18F]fluoro-2-deoxy-D-glucose (FDG) to simulate high- and low-radioactivity conditions corresponding to brain FDG PET and amyloid PET acquisitions, respectively. Two separate phantom measurements were performed accordingly. Motion simulation was conducted using a custom-designed motion generator, incorporating 15° and 30° rotations about the z-axis, 3° and 6° rotations about the x-axis, and 5 mm and 10 mm translations along the z-axis in the PET image coordinates. The data-driven MC was applied with frame durations of 1, 2, 5, 10, and 20 s for motion estimation. The estimated motions were compared with the motions measured using an external optical tracker system. %contrast and gray matter coefficient of variation (CV%) were calculated from the motion-corrected PET images.

RESULTS

The motion generator system successfully reproduced the designed motions. Motion estimation remained stable under high-radioactivity condition but showed reduced stability under low-radioactivity condition, particularly with shorter frame durations. Under both conditions, longer frame durations led to underestimation of continuous motion. The data-driven MC improved %contrast and gray matter CV% across all conditions, with shorter frame durations providing better correction for quick or continuous motions. However, shorter frame durations increased statistical noise, especially under low-radioactivity condition.

CONCLUSION

The data-driven MC effectively improved the quality of motion-affected PET images under both high- and low-radioactivity conditions, indicating its broad applicability. However, correction accuracy deteriorated under the lower-radioactivity condition.

Для цитирования: Kamitaka Yuto, Sakata Muneyuki, Oda Keiichi, Katsuki Akie, Kawakami Hirofumi, Wagatsuma Kei, Kobayashi Masato, Ishii Kenji. Quantitative validation of data-driven motion correction for brain PET using phantom with motion generator system.. EJNMMI physics. 2025. DOI: 10.1186/s40658-025-00820-w.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.