MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jelectrocard.2018.05.010

Медицинская библиотека · Journal of electrocardiology · 2018

T-wave oversensing is more dependent on ICD sensitivity algorithm than lead sensing configuration.

Punjabi Anil H, Dermovsesian Arlene, Karnik Ankur A

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

Modern implantable cardioverter defibrillators (ICD) employ dynamic sensing algorithms in order to protect against fine ventricular fibrillation without oversensing intrinsic activity. We present a patient with a Medtronic ICD who had inhibition of pacing and not inappropriate shocks due to T wave oversensing (TWOS) in both true bipolar (TB) and integrated bipolar (IB) sensing configurations. Rather than alternatives such as lead revision or programming to an unacceptably insensitive value, this was solved by exchanging for a Boston Scientific ICD. Although the literature suggests lead sensing configuration impacts TWOS, this case demonstrates ICD sensitivity algorithm may be a key determinant.

Для цитирования: Punjabi Anil H, Dermovsesian Arlene, Karnik Ankur A. T-wave oversensing is more dependent on ICD sensitivity algorithm than lead sensing configuration.. Journal of electrocardiology. 2018. DOI: 10.1016/j.jelectrocard.2018.05.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.