Медицинская библиотека · Nuklearmedizin. Nuclear medicine · 1997

[How great is radiation exposure in the area surrounding patients after administration ov 99mTc-sestamibi?].

Kurtaran A, Preitfellner J, Schaffarich P, Scheuba C, Niederle B, Virgolini I, Havlik E

СтатьяПеревод пока не подготовлен

Аннотация

AIM: The aim of the present study was to estimate the additional radiation exposure to personnel, other patients and members of the family caused by patients who had been injected with 99mTc-Sestamibi (Cardiolite DuPONT PHARMA) for preoperative localization of parathyroid adenoma.

METHODS

Dose rates were measured from 16 patients who had received an intravenous injection of 600 +/- 50 MBq 99mTc-Sestamibi. All measurements were performed with a portable dosimeter (Berthold LB 133) at 3 different distances (0.5, 1 and 2 m) at 10 min, 180-200 min and 24 h after administration of the tracer.

RESULTS

The dose rates amounted to 20.5 microSv/h at 0.5 m, 5.25 microSv/h at 1 m and 1.55 microSv/h at 2 m distance from patients respectively. The biological half-life was 54 h. The calculated maximal dose to other persons in the waiting area was 31.2 microSv, to family members 27.6 microSv and to nurses in a ward 31.2 microSv.

CONCLUSION

Our results indicate that the calculated maximal radiation exposure for personnel, family members and other patients even under very unfavourable conditions was below the maximal allowed dose for non-professionally exposed persons.

Для цитирования: Kurtaran A, Preitfellner J, Schaffarich P, Scheuba C, Niederle B, Virgolini I, Havlik E. [How great is radiation exposure in the area surrounding patients after administration ov 99mTc-sestamibi?].. Nuklearmedizin. Nuclear medicine. 1997.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.