Assessment of the biomechanical properties of the cornea with the ocular response analyzer in normal and keratoconic eyes.

Shah Sunil, Laiquzzaman Mohammed, Bhojwani Rajan, Mantry Sanjay, Cunliffe Ian

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare hysteresis, a novel measure of ocular rigidity (viscoelasticity) in normal and keratoconic eyes.

METHODS

The study consisted of 207 normal and 93 keratoconic eyes. Eyes were diagnosed as keratoconic based on clinical examination and corneal topography. The hysteresis was measured by the Ocular Response Analyzer (ORA; Reichert Ophthalmic Instruments, Buffalo, NY). The data were recorded by Generation 3 software for the ORA. Central corneal thickness (CCT) was measured with a handheld ultrasonic pachymeter in the midpupillary axis.

RESULTS

The mean hysteresis was 10.7 +/- 2.0 (SD) mm Hg (range, 6.1-17.6) in normal eyes compared with 9.6 +/- 2.2 mm Hg (range, 4.7-16.7) in keratoconic eyes. The difference was statistically significant (P < 0.0001, unpaired t-test). Mean CCT in the normal and keratoconic eyes was 545.0 +/- 36.4 microm (range, 471-650) and 491.8 +/- 54.7 microm (range, 341-611), respectively; the difference was significant (P < 0.0001, unpaired t-test).

CONCLUSIONS

Hysteresis was significantly higher in normal than in keratoconic eyes. It may be a useful measurement in addition to CCT, when assessing ocular rigidity, and may be of particular importance when trying to correct intraocular measurements for increased or decreased ocular rigidity. Long-term studies of change in hysteresis may provide information on the progression of keratoconus.

Для цитирования: Shah Sunil, Laiquzzaman Mohammed, Bhojwani Rajan, Mantry Sanjay, Cunliffe Ian. Assessment of the biomechanical properties of the cornea with the ocular response analyzer in normal and keratoconic eyes.. Investigative ophthalmology & visual science. 2007. DOI: 10.1167/iovs.04-0694.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.