DOI 10.1016/j.jcrs.2006.05.030
Optical coherence tomography to assess intrastromal corneal ring segment depth in keratoconic eyes.
Аннотация
PURPOSE: To investigate intrastromal corneal ring segment depth with a high-speed corneal optical coherence tomography (OCT) system.
SETTING
Doheny Eye Institute, University of Southern California, Los Angeles, California, USA.
METHODS
A prospective observational case series comprised 4 eyes of 4 patients receiving Intacs intrastromal corneal ring segments (Addition Technology, Inc.) for keratoconus. Optical coherence tomography (OCT) was performed between 7 days and 43 days after implantation.
RESULTS
The slitlamp impression of intrastromal corneal ring segment implantation depth did not correlate well with OCT measurements (r(2) = 0.68). The fractional implantation depth was correlated with several surgical variables using a stepwise multivariate regression model, and 2 statistically significant correlations were found. The position of the distal portions of the ring segments was shallower than that of the portion closer to the insertion site (P = .003). Segments placed in the inferior cornea (P = .008) experienced more distal shallowing. Shallower depth was associated with greater fractional anterior stromal compression (P = .04).
CONCLUSIONS
Shallower placement of intrastromal corneal ring segments may result in more complications, such as epithelial-stromal breakdown and extrusion, because of the greater anterior stromal tensile strain. The distal and inferior portions of intrastromal corneal ring segments tended to be placed at a shallower depth. Optical coherence tomography provided precise measurement of ring segment depth and may help identify implants that pose a greater risk for depth-related complications.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.