MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3928/1081-597x-20061001-06

Медицинская библиотека · Journal of refractive surgery (Thorofare, N.J. : 1995) · 2006

INTACS inserts using the femtosecond laser compared to the mechanical spreader in the treatment of keratoconus.

Rabinowitz Yaron S, Li Xiaohui, Ignacio Teresa S, Maguen Ezra

Клинический случайСравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To determine the efficacy of INTACS insertion using a femtosecond laser in the treatment of keratoconus and to compare it to the technique using a mechanical spreader.

METHODS

INTACS were inserted in 10 eyes using the mechanical spreader to create the channels and subsequently on another 20 eyes using the femtosecond laser. Uncorrected (UCVA) and best spectacle-corrected visual acuity (BSCVA), manifest refraction, and corneal topography were measured prior to surgery, at 6 months (femtosecond group), and 1 year (mechanical group). Pre- and postoperative data were analyzed to determine changes in the above parameters.

RESULTS

Both groups showed significant reduction in average keratometry (K), spherical equivalent refraction, BSCVA, UCVA, surface regularity index (SRI), and surface asymmetry index (SAI). The laser group performed better in all parameters except change in SRI. Results of the laser versus the mechanical spreader were as follows: reduction in spherical equivalent refraction (3.98 vs 2.96), change in average K (2.91 vs 2.52), improvement in UCVA (4.13 vs 3.63), improvement in BSCVA (3.92 vs 1.63), change in SRI (0.37 vs 0.64), and change in SAI (1.00 vs 0.70). Statistical analysis, however, did not reveal any statistically significant differences between the two groups for any single parameter studied. The biggest improvement in the laser group versus the mechanical group was BSCVA (P=.09). Overall success, defined as contact lens or spectacles tolerance, was 85% in the laser group and 70% in the mechanical group.

CONCLUSIONS

Inserting INTACS using the femtosecond laser to create the channels is as effective as using the mechanical spreader.

Для цитирования: Rabinowitz Yaron S, Li Xiaohui, Ignacio Teresa S, Maguen Ezra. INTACS inserts using the femtosecond laser compared to the mechanical spreader in the treatment of keratoconus.. Journal of refractive surgery (Thorofare, N.J. : 1995). 2006. DOI: 10.3928/1081-597x-20061001-06.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.