Hospital-based health technology assessment of central dialysis fluid delivery system for hemodialysis patients.

Cheng Wendi, Wang Haiyin, Yuan Jiangzi, Li Ying, Luo Yashuang, Fu Yuyan, Gu Leyi, Xue Jun, Jin Chunlin, Zheng Yin

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

This study employed a hospital-based health technology assessment (HB-HTA) to evaluate the economic value of the central dialysis fluid delivery system (CDDS) compared to the single-patient dialysis fluid delivery system (SPDDS) in hemodialysis. The findings offer evidence to support hospital procurement and clinical implementation decisions regarding CDDS.

METHODS

A comparative cost analysis was conducted using micro-costing data collected from three hospitals. Costs were estimated for both CDDS and SPDDS, and economic efficiency was evaluated across multiple hemodialysis machine configuration scales. The analysis included total costs, cost structures, and cost savings, along with their constituent components, to identify scenarios in which CDDS provides economic advantages.

RESULTS

In a dialysis center with 6 hemodialysis machines operating two shifts per day and serving 12 patients daily, CDDS reduced costs by 0.05% compared to SPDDS. As the number of machines increased from 6 to 50, the cost reduction rate rose from 0.05% to 21.08%, demonstrating enhanced economic benefits at larger scales. In a 50-machine center, CDDS saved approximately RMB 84 per treatment session and RMB 2,601,108 annually. The majority of cost savings came from reductions in consumables (74%) and labor (24%). One-way sensitivity analysis confirmed that CDDS consistently yielded cost savings, supporting the robustness of the costing model.

CONCLUSION

CDDS demonstrates significant economic value over SPDDS by reducing consumable and labor costs, proving to be a cost-effective solution in dialysis centers with 6 or more machines serving at least 12 patients daily. Clinical engineers play a key role in its implementation. CDDS is suitable for promotion in medium- to large-scale dialysis centers in China.

Для цитирования: Cheng Wendi, Wang Haiyin, Yuan Jiangzi, Li Ying, Luo Yashuang, Fu Yuyan, Gu Leyi, Xue Jun, Jin Chunlin, Zheng Yin. Hospital-based health technology assessment of central dialysis fluid delivery system for hemodialysis patients.. BMC nephrology. 2025. DOI: 10.1186/s12882-025-04484-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.